3 setting the dc output voltage – Metrohm 731 Relay Box User Manual

Page 15

Advertising
background image

2.3 Setting the DC output voltage

Relay Box 731, Instructions for use

9

2.3 Setting the DC output voltage

The common output voltage for the direct current outputs (DC1 and
DC2) is set inside the Relay Box housing.

Switch off the instrument and pull out the line plug. Loosen the
four knurled screws of the protective cover and remove them.

Right-hand side
of housing with-
out protective
cover

Rear view of
housing without
protective cover

The screws marked with an arrow in the diagram must be loos-
ened.

Carefully remove the instrument PCB from the housing.

To set the output voltage the plug-in bridge P9 on the relay box
motherboard must be fitted with the appropriate jumper .

The location of plug-in bridge P9 can be seen in the diagram of the
instrument motherboard shown below.

+18V

+10V

+24V

0V*
+5V

1

12

P9

6

7

Set the jumper, which is located in the +24 V position
as standard, to the position corresponding to the re-
quired output voltage with the help of tweezers or ta-
per-nose pliers.

+18V

+10V

+24V

0V*

+5V

1

12

P9

6

7

Carefully reinsert the motherboard in the housing and screw the in-
strument together. Use the adhesive labels provided to note the set
output voltage on the rear panel of the instrument.

Jumper

Advertising