7 injection valve, 8 suppressor, 9 metrohm co2 suppressor (mcs) – Metrohm 930 Compact IC Flex SeS User Manual
Page 101: Injection valve, Suppressor, Metrohm co, Suppressor (mcs), 9 metrohm co
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
93
Safety shutdown
Function
Automatic shutdown upon reaching the pressure limit values
Maximum pres-
sure limit
■
Adjustable from 0.1 - 50 MPa (1 - 500 bar)
■
The pump is automatically shut down at the first piston stroke
above the maximum limit value
Minimum pres-
sure limit
■
Adjustable from 0 - 49 MPa (0 - 490 bar)
■
The shutdown mechanism is inactive at 0 MPa
■
The shutdown mechanism becomes active 2 minutes after system
start
■
The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit
7.7
Injection valve
Actuator switch-
ing time
typ. 100 ms
Maximum operat-
ing pressure
35 MPa (350 bar)
Material
PEEK
7.8
Suppressor
Resistance to sol-
vents
No restriction
Switching time
typ. 100 ms
7.9
Metrohm CO
2
Suppressor (MCS)
Material
Fluoropolymer
Resistance to sol-
vents
No restriction (except PFC)
Underpressure
Working range
Microprocessor controlled/stabilized
Time to estab-
lish after start-
ing
< 30 s
Capillary volume
400 µL
Recommended
flow range
0.1 - 1.0 mL/min