7 injection valve, 8 suppressor, 9 metrohm co2 suppressor (mcs) – Metrohm 930 Compact IC Flex SeS User Manual

Page 101: Injection valve, Suppressor, Metrohm co, Suppressor (mcs), 9 metrohm co

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Technical specifications

930 Compact IC Flex SeS (2.930.1400)

■■■■■■■■

93

Safety shutdown

Function

Automatic shutdown upon reaching the pressure limit values

Maximum pres-
sure limit

Adjustable from 0.1 - 50 MPa (1 - 500 bar)

The pump is automatically shut down at the first piston stroke
above the maximum limit value

Minimum pres-
sure limit

Adjustable from 0 - 49 MPa (0 - 490 bar)

The shutdown mechanism is inactive at 0 MPa

The shutdown mechanism becomes active 2 minutes after system
start

The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit

7.7

Injection valve

Actuator switch-
ing time

typ. 100 ms

Maximum operat-
ing pressure

35 MPa (350 bar)

Material

PEEK

7.8

Suppressor

Resistance to sol-
vents

No restriction

Switching time

typ. 100 ms

7.9

Metrohm CO

2

Suppressor (MCS)

Material

Fluoropolymer

Resistance to sol-
vents

No restriction (except PFC)

Underpressure

Working range

Microprocessor controlled/stabilized

Time to estab-
lish after start-
ing

< 30 s

Capillary volume

400 µL

Recommended
flow range

0.1 - 1.0 mL/min

Advertising