Metrohm 787 KF Titrino User Manual

Page 44

Advertising
background image

3.1. Error and special messages

787 KF Titrino

40

Error messages in connection with the data transfer

If neither a computer nor a printer is attached, the report output at the end of the titration

must be switched off:



Receive errors:

E36

Parity

Exit: <QUIT> and set corresponding quantity the

same on both instruments

E37

Stop

bit

Exit: <QUIT> and set corresponding quantity the

same on both instruments

E38

Overrun error. At least 1 character could not be read.

Exit: <QUIT>

E39

Overflow of the receive buffer of the Titrino (> 82

characters).

Exit: <QUIT>




Send errors:

E40

DSR=OFF

E41

DCD=OFF

E42

CTS=OFF

Handshake unsatisfactory for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready

to receive?

E43

The transmission of the Titrino has been interrupted

with XOFF for at least 3 s.

Exit: <QUIT>.

E44

The RS interface parameters are no longer the same

for both instruments.