3.1. Error and special messages
787 KF Titrino
40
Error messages in connection with the data transfer
If neither a computer nor a printer is attached, the report output at the end of the titration
must be switched off:
Receive errors:
E36
Parity
Exit: <QUIT> and set corresponding quantity the
same on both instruments
E37
Stop
bit
Exit: <QUIT> and set corresponding quantity the
same on both instruments
E38
Overrun error. At least 1 character could not be read.
Exit: <QUIT>
E39
Overflow of the receive buffer of the Titrino (> 82
characters).
Exit: <QUIT>
Send errors:
E40
DSR=OFF
E41
DCD=OFF
E42
CTS=OFF
Handshake unsatisfactory for more than 1 s.
Exit: <QUIT> Is the receiver switched on and ready
to receive?
E43
The transmission of the Titrino has been interrupted
with XOFF for at least 3 s.
Exit: <QUIT>.
E44
The RS interface parameters are no longer the same
for both instruments.