5 analogue output test, 6 motor timer test – Metrohm 787 KF Titrino User Manual
Page 49
3.3.
Diagnosis
787 KF Titrino
45
• Press <0>.
date/time
cylinder code
• Press <Enter>.
date XX-XX-XX
XX:XX:XX
code:D0 XX ml
activated dosing unit
mL-code
• Check date and time.
• Check whether the ml-code does correspond with the exchange unit placed or not.
For the sake of completeness, different exchange units can be inserted to check their ml code.
If desired, the ex-
change unit can be removed again. If no exchange unit is inserted display does not show the ml-code but „check
exchange unit !“.
• Press <Clear>.
diagnose press key 0...9
5 Analogue output test
A voltage can be set at the analog output (sockets at D) using the keypad. But this should not exceed
±
2000 mV. This volt-
age can also be used for the calibration of a connected recorder.
• Connect a voltage measuring instrument (voltmeter, DVM, recorder) to the ana-
logue output
(10).
• Press <3>.
analog output-1 test
• Press <Enter>.
analog output-1 test
V-out = XXmV
Enter a voltage value in the range
±
2000 mV using the keypad. After the <ENTER> key has
been pressed, this value appears as a voltage at the analog output..
Read off value an the connected voltmeter and compare with the mV value on the display.
(Tolerance
±
2 mV)
• Press <Quit>.
• Disconnect voltmeter.
6 Motor timer test
• Press <6>.
motor-timer test
• Press <Enter>.