Suspension array system validation report – Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 22

Advertising
background image

your product insert for values specific to your product control number..

Parameter

Specification

Measured Value

Pass/Fail

DD Median

4774 - 6593

6133

Pass

CL1 Median

3383 - 4135

3615

Pass

CL1 CV%

2.00 - 7.00

3.52%

Pass

CL2 Median

3520 - 4302

3883

Pass

CL2 CV%

3.00 - 8.00

6.93%

Pass

RP1 Median

3509 - 4290

3878

Pass

RP1 CV%

4.00 - 10.00

8.89%

Pass

Parameter

Specification

Measured Value

Pass/Fail

% Carryover

< or = 4.0%

1.3%

Pass

A. Calculated Values

B. Raw Values

Parameter

Specification

Measured Value

Pass/Fail

Bead

Median FI

Dynamic Range

4.33 - 4.45

4.40

Pass

Blank

2

Linearity

> 0.995

1.000

Pass

1

8

Slope 33.94

-

39.42

37.57

Pass

2

54

Accuracy

> 90.00%

96.94%

Pass

3

656

Threshold

< 6 MFI

2

Pass

4

1576

5

4453

6

25059

A. Classify Efficiency

Classify Bead

Specification

Measured Value

Pass/Fail

Bead A (1)

> 80.0%

98.1%

Pass

Bead B (4)

> 80.0%

93.5%

Pass

Bead C (40)

> 80.0%

90.5%

Pass

Bead D (54)

> 80.0%

95.3%

Pass

Bead E (100)

> 80.0%

93.0%

Pass

B. DD Efficiency

Parameter

Specification

Measured Value

Pass/Fail

DD Efficiency

> or = 75.0%

96.7%

Pass

Comments:

Reviewed by:

Date:

III. Reporter Validation

IV. Classify Validation

II. Fluidics Validation

Test Performed by: JPocekay

Date/Time: 04-Aug-04 11:55 AM

Expiration Date: 15-Oct-06

Reader Serial #: LX10000266002

Access Level: Unrestricted

Validation Kit Control #: 999999 (Low)

RP1 Target Value: 3885

RP1 PMT Voltage: 571.67

Result: Passed

Result: Passed

Result: Passed

Result: Passed

I.Optics Validation

Suspension Array System Validation Report

20

4110185B.qxp 9/25/2007 11:55 AM Page 20