Acquire Data
1. Shake the assay plate at 900 rpm for 30 sec immediately before

acquiring data. Failure to do so will result in increased read time due

to bead settling.

2. Visually inspect the plate and ensure that the assay wells are filled with

buffer prior to placing the plate in the Bio-Plex microplate platform.

3. Slowly remove the sealing tape and any plate cover before placing

the plate in the reader.

4. Select the appropriate protocol (.pbx) file.

5. Select Start and save the report (.rbx) file.

6. Begin data acquisition.

NOTE: If acquiring data from more than one plate, empty the waste

bottle and refill the sheath bottle after each plate (if HTF not present).

Select Wash Between Plates

and follow the instructions for

fluidics maintenance. Then repeat the Prepare Protocol and
Acquire Data steps. We recommend using the Wash Between

Plates command after every plate run to reduce the possibility of

clogging the instrument.

7. When data acquisition is complete, select Shut Down

and follow

the instructions.

Reacquire Data
It is possible to acquire data from a well or plate a second time using the

Rerun/Recovery mode located below Start in Step 7 (Run Protocol). For

details, refer to the Bio-Plex Manager instruction manual.