Reaquire Data
It is possible to acquire data from a well or plate a second time using the

Rerun/Recovery mode located below Start in Step 7 (Run Protocol). For

details, refer to the Bio-Plex Manager instruction manual.

Bio-Plex Manager Software Version 4.0

Bio-Plex Manager version 4.0 does

not include the 25-bead map. Hence,

the 100-bead map is used. Follow the instructions above for Bio-Plex

Manager version 4.1 and 4.1.1 for preparing and calibrating the system.

Be careful to calibrate on Low RP1 and in Step 7 (Run Protocol) under

Advanced Settings, set the

DD gates to 5,000 (low) and 32,000 (high).

Acquire Data
1. Shake the assay plate at 1,100 rpm for 30 sec immediately before

acquiring data. Failure to do so will result in increased data

acquisition time due to bead settling.

2. Visually inspect the plate and ensure that the assay wells are filled with

buffer prior to placing the plate in the Bio-Plex microplate platform.

3. Slowly remove the sealing tape and any plate cover before placing

the plate in the reader.

4. Select Step 7 (Run Protocol):

a)

Choose

100 beads per region.

b)

In Advanced Settings, set the Bead Map to

100 region.

c)

In Advanced Settings, set the sample size to

50 µl.

d)

In Advanced Settings, confirm that the DD gates are set to

5,000

(low) and

32,000 (high).

NOTE: These settings are already selected in preset downloaded

protocols. If creating a new protocol, manually change the default

setting to reset the gates to 5,000 (low) and 32,000 (high).

e)

Select Start and save the report (.rbx) file. Then follow the

instructions for data acquisition.