System Settings
User Guide
| 69
System Settings
The System Settings dialog box enables you to customize your system. From this
dialog box you can
Set the delay volume
Control the flow rate to prevent overpressure
Enable remote access to ChromLab from an iPad, Android device, or
another PC
Reset the trace settings, the minimum and maximum y-scale values, and
the trace colors in the chromatogram to their factory default settings
Enable or disable external detectors that are connected to the NGC
instrument via a SIM device
Enable or disable the ability to send UV signals to external devices that are
connected to the NGC instrument via a SIM device
Set air sensors to detect either end of sample or end of buffer
Enable the system to send email messages about system events to a list of
users
Set a unique name for the NGC system
Note:
This dialog box is accessible in manual mode. The settings are saved
and the fraction collector is also synchronized in method runs.
Delay Volume Tab
Delay volume is the volume of plumbing between the UV detector and the fraction
collector drophead and/or the outlet valve. It indicates the delay between the
detection of the fraction’s UV peak and the time it takes for this peak to traverse the
volume of tubing and flow cells to the fraction tube into which it is collected.