4.0 SYSTEM PLUMBING
This chapter begins with discussion of plumbing practices and general guidelines, followed by a general pro-
cedure for plumbing the major system components. Later sections discuss how to plumb Bio-Rad’s low
pressure valves.
4.1 GENERAL GUIDELINES FOR PLUMBING THE SYSTEM
The BioLogic LP uses the following tubing:
•
System tubing: 0.8mm ID or 1.6mm ID Tygon tubing.
•
Pump Tubing: 0.8mm ID, 1.6mm ID, or 3.2mm ID PharMed tubing. Refer to the following table
when selecting a tubing.
Warning
Do not use pump tubing with wall thickness greater than 1.0 mm. Using tubing with a wall thickness
greater than 1.0 mm can damage the pump and void the warranty.
Table 4
Comparison of Flow Rate Ranges for Different Tubing IDs
22
System Plumbing
1.0
.02
0.8/0.8 mm
0.8 mm: 0.05 to 1.75 ml/min
1.6 mm: 0.1 to 6.5 ml/min
3.2 mm: 0.2 to 20.0 ml/min
FLOW RATE RANGE ml/hr
FLOW RATE RANGE ml/min
OPTIMAL RANGE
MAXIMUM RANGE:
TUBING SIZE
(ID/WALL)
1.6/0.8 mm
3.2/0.8 mm
.04
.06 .08
.2
.4
.6 .8
2
4
6 8
20
3.0
5.0
10
30
50
100
300 500
1000