Cartridge Replacement

• Always use protective gloves when removing the cartridges.

• Insert the metal pin in the pin insertion point hole on the

control mechanism.

• Hold the control mechanism at the pin position and not at

the on/off trigger. This helps grip the body without
damaging the trigger.

• Unscrew the cartridge from the control mechanism in a

counter clockwise direction.

• Remove the pin.

• Inspect the control mechanism for o-ring. If o-ring is

missing, do not attach the cartridge. Contact Integra
Miltex for o-ring replacement.

• Do not remove cartridge until empty.

• Any remaining gas will cause it to be more difficult to

remove the cartridge.

CryoSolutions

®

Treatment

Contact Applications

• Always use protective gloves when using CryoSolutions

®

.

• Remove the plastic protective cap.

• Place the Dispensing tip perpendicular to and directly on the

skin surface and area to be treated.

• Do not apply additional pressure on the skin surface and area

to be treated.

• Depress the trigger to open the valve and immediately

liquefied gas will be delivered to the dispensing tip and on to
the skin.

• Application times are normally very brief as CryoSolutions

®

provides high performance cold temperature to the
treatment area (exceptions may include plantar warts).

• Most lesions are removed with a single treatment. However,

others may require a second treatment or even further
applications, such as plantar warts.