Salter Brecknell SBI-505 User Manual
Page 10
8
[0080.00]
[CAL End]
As bellows:
Input the 0080.00, count down ,
then indicator shows 0080.00 ,
calibration is over.
If you want to set application
function parameter. Press
“PRINT” if you want to exit
press “Count”
8
press
press
press
or
[C07 ]
[07 0]
[07 1]
Default parameters setting
option:0=non-restore default
parameters
1=restore default parameters
Note: after the above
parameters setting finish, please
do not select 1 to avoid losing
new setting parameters.
Application function parameters setting chart
Function
Setting
Item
parameters setting and instruction
key tone
C08
warning
tone
Options: 0 = no key tone
1 = key tone
Automatic
power off
C09
Automatic
power off
option
:0=close auto power off
10= power off automatically if no
change within 10 minute.
30= power off automatically if no
change within 30 minute.
60= power off automatically if no
change within 60 minute.
Power saving
setting
C10
Power saving
setting
LED Version ONLY:
option
:0= close power saving setting