Maintenance – Salter Brecknell SBI-505 User Manual

Page 20

Advertising
background image

18

6. Maintenance

6.1 Regular error and solution

ERROR REASON

SOLUTION

UUUUUU

1. Overload
2. wrong connection
with load cell
3. Load cell has
problem.

1. reduce the weight
2. check load cell connection
3. Inspect load cell. Check the
input and output voltage 0R
input and output resistance to
judge it is good or not.

nnnnnnn

1. Under load
2. LC connection
3. load cell has
problem

1. Check platform if it is level or

not.

2. Check load cell connection.

3. Check load cell

:Check the

input and output voltage 0R
input and output resistance to
judge it is good or not.

ERR1


During calibration, no
input of weights or the
weight is overload

Input the correct weights

ERR2


During calibration , the
weights is below than
Min. required weights

The calibration weights
Minimum is 10% of Max. cap.
Recommend 60%-80% of Max.
Cap.