Brookfield DV-I Prime User Manual

Page 44

Advertising
background image

Brookfield Engineering Laboratories, Inc.

Page 44

Manual No. M/07-022-D0613

A.3 Setting the Gap

1. Move the toggle switch to the right; this will

turn on (enable) the Gap Setting Feature. The

Pilot (red) light will be illuminated.

2. If the contact light (yellow) is illuminated, turn

the micrometer adjustment ring clockwise (as

you look down on the instrument) until the

light is just breaking contact, i.e., flickering

(see Figure A-5).

3. If the yellow contact light is not illuminated,

slowly

turn the micrometer adjustment ring in

small increments (one or two scale divisions)

counter-clockwise.

Continue moving the micrometer adjustment

r i n g

s l o w l y

counter-clockwise until the contact light

(yellow) turns on. Back off (rotate clockwise)

until the light is just breaking contact, i.e.,

flickering.

4. Adjust the sliding reference marker, right or

left, to the closest full scale division mark (see
Figure A-6)
.

5. Turn the micrometer adjustment ring one scale

division to the left to meet the line on the

sliding reference marker. THE YELLOW
CONTACT LIGHT SHOULD GO OFF
.

6. You have established the gap space needed for

measurement. NOW TURN THE TOGGLE
SWITCH OFF (LEFT); THE RED PILOT
LIGHT SHOULD GO OFF.

7. Carefully remove the sample cup.

Moves Away

from Hit Point

(clockwise)

LEFT

x

Moves Towards

Hit Point

(counter-clockwise)

RIGHT

Figure A-5

Figure A-6

This viscoity of “electrically conductive”

fluids may be affected if readings are taken

with the Electronic Gap Setting feature

“on”. Be sure to shut the feature “off”

before taken readings!

Notes:

1. The cup may be removed and replaced without resetting the gap, if the micrometer

adjustment ring has not been moved.

2. Remove the spindle from the viscometer when cleaning.

3. Re-establish the hit point every time the spindle is attached/detached.

Advertising