Page 21
Brookfield Engineering Labs., Inc.
page 21
Manual No. M14-223
II.5.2 Electrical Connections RST-CPS
Connections for the electrical components of the RST-CPS Rheometer are located on the back of the
instrument.
Figure II-8
Connecting elements on the back of the instrument in Figure II-8:
1 LAN (network)
4 RS 232 port
2 PC (USB)
5 USB stick or printer
3 ON / OFF button
6 ON / OFF control light
Caution!
Connecting the measuring instrument
Connect or disconnect any cables from and to the RST-CPS Rheometer only with the
instrument turned off.
Be sure that your USB Flash Drive is formatted in FAT32. Otherwise it doesn’t work.
II.5.3 Temperature Sensor RST-CPS
The temperature sensor Pt100 is inside the bottom measuring plate. The temperature is measured
continuously and is displayed in the Touch Screen Display alternately with the Date.
II.5.4 AC Adapter RST-CPS
The AC adapter is used to power the RST-CPS Rheometer. Do not use any other device to supply the
RST-CPS Rheometer with power than the AC adapter of IP class 4/2 delivered by Brookfield.
Danger!
AC adapter
Always connect the AC adapter to a properly grounded socket. To avoid electric shock or
damage to system components, always use a properly grounded plug to connect the AC
adapter.
Connecting the AC adapter:
• Insert the connecting socket of the power cord into the AC adapter until the plugs locks into place.