KEYENCE LK-H2 User Manual
Page 67
A-3
A
During configuration
Head settings
OUT settings
Error message
Cause
Correction
ABLE tuning
Setting range is 1 to 99.
The ABLE field contains a blank
or is set to 0.
Set the ABLE field with a value
from 1 to 99.
Set so the lower limit < = upper
limit.
The lower limit for ABLE is set to a
value greater than the upper limit.
Set with a value so that the lower
limit < = upper limit.
Alarm cycles to hold
Illegal input. Input a value from 0
to 9999.
The alarm cycles to hold field
contains a blank.
Set the alarm cycles to hold field
with a value from 0 to 9999.
Error message
Cause
Correction
Scaling input
Illegal input.
Input a value from (-99.9999 to
99.9999)*.
The scaling setting field contains
a blank, non-numeric value, or a
value outside the setting range.
Set the scaling setting input value
to a value within the displayed
range.
Input value 1 and input value 2
cannot be set to the same value.
The input value 1 and input value
2 fields for the scaling setting are
set to the same value.
Set the input value 1 and input
value 2 fields for the scaling
setting to a different value.
Scaling display
Illegal input.
Input a value from (-99.9999 to
99.9999) *.
The scaling setting for displaying
field contains a blank, non-
numeric value, or a value outside
the setting range.
Set the scaling setting display
value to a value within the
displayed range.
Scaling in general
Set so that |(display value 2 -
display value 1) / (input value 2 -
input value 1) | <= 2.
The slope of the scaling setting
exceeds 2.
Set the input and output values
so that the slope of the scaling
setting is less than or equal to 2.
Offset
Illegal input.
Input a value from (-99.9999 to
99.9999) *.
The offset field contains a blank,
non-numeric value, or a value
outside the setting range.
Set the offset field to a value in
the displayed range.
Tolerance upper/lower limit
Illegal input.
Input a value from (-99.9999 to
99.9999) *.
The tolerance upper/lower limit
fields contain a blank, non-
numeric value, or a value outside
the setting range.
Set the tolerance upper/lower
limit fields to a value in the
displayed range.