➨ Set hysteresis Δp by using keys Δp▲ or Δp▼:

☞ Set mode is activated by pressing key shortly: Set Δp ap-

pears.

☞ Short tips toggle in steps of 1 mbar.

☞ Continuous pressing activates ramp: Hysteresis Δp changes

with increasing speed.

Continuous pumping

Pumping down until the ultimate vacuum of the pump is
attained
(e. g. pump down to dry off volatile liquids in the sys-
tem):
If the controller is in mode two-point control:
➨ Press key MODE to change to continuous pumping.

☞ Changing the mode is only possible if the process control is
inactive.
☞ The two-point control display disappears.

MODE

Start: Press key START/STOP.

☞ Arrow down and valve and/or pump symbol are flashing.

Stop: Press key START/STOP.

➨ Selecting the set point by using keys p▲ or p▼.

☞ Set mode is activated by pressing key shortly: Set p ap-

pears.

☞ Short tips toggle p in steps of 1 mbar.

☞ Continuous pressing activates ramp: set point p changes

with increasing speed.

☞ Changing set point p leads at the same time to an automatic

preselection of hysteresis

Δp.

☞ Hysteresis Δp can be changed manually at any time (similar

to changing set point p) by using keys

Δp, see ”Manual set-

ting of hysteresis

Δp“.

Two-point control

Manual setting of set point p (boiling point)

➨ Start continuous pumping.

☞ Switch over to continuous pumping by pressing key MODE

if controller is in two-point control mode.

☞ Changing the mode is only possible if process control is

inactive. Display two-point control disappears.

MODE

Semiautomatic determination of set point p (boiling point)

START

STOP

p

267

mbar

ΔΔΔΔΔp

67

mbar

Manual setting of hysteresis

ΔΔΔΔΔp