Attention: Perform a leak test always after opening the pump by
using an appropriate leak detector (e. g. helium leak detector)!
Ensure that the diaphragm clamping disc never hits the head
cover while pump is running! Noise check!
If the pump does not achieve the ultimate pressure:
☞
In case the diaphragms and valves have been replaced, a run-in period of several
hours is required before the pump achieves its ultimate vacuum.
If all pump heads achieve a vacuum below 120 mbar but pump does not achieve the
ultimate total pressure:
Check hose connectors between pump heads and manifolds for leaks. If necessary
recheck pump chamber.