Interface parameters

The CVC 3000 controller is equipped with a serial interface (RS 232C,

nine-pin Sub-D-plug).
+

Plug-in or remove the cable (cable RS 232C) from the interface only if

the equipment is switched off.

+

The interface is

not electrically isolated from the measuring circuit.

+

For optimal electromagnetic compatibility assemble an interface filter

(cat. no.: 638235).

The controller is fully operable via the serial interface. Measuring results,

preselections and the status of the controller can be read at any time.

The factory-set read and write commands are completely compatible

with the VACUUBRAND CVC 2000 controller (see sections ”Read / Write

commands CVC 2000”). An extended instruction set is available using the

command ”CVC 3” (see sections ”Read / Write commands CVC 3000”).

Setting of the interface

Set the interface parameters directly at the CVC 3000 controller. The fac-

tory set values are underlined.

Edit and confirm the interface parameters in the ”Configuration” menu in

”RS-232” submenu using the selection knob.

Baud: 2400, 4800, 9600 or 19200

Parity: 8-N-1, 7-O-1 or 7-E-1

Handshake: Off, Xon-Xoff or RTS-CTS

Remote: On or Off

Timeout: Sending 1s, receiving 10s.

In remote mode (”Remote On”, with the ”PC symbol” in the display) all

keys at the controller are inoperable.

To return to the manual operation of the controller, set the control-

ler to ”Remote off” in menu configuration: Switch off the controller.

Then switch the controller back on, and press the selection knob

within 2s.