Automatic identification of connected components – VACUUBRAND DCP 3000 User Manual

Page 32

Advertising
background image

page 32 of 64

Automatic identification of connected components

When switching on the device, the configuration of the connected com-

ponents is checked.

Connected components (valves, pressure transducers, etc.) are au-

tomatically identified, used and monitored until the device is switched

off. Reconfiguration is possible by switching the DCP 3000 off/on. Identi-

cal components must be configured beforehand to different VACUU•BUS

addresses (e.g., VSK 1 and VSK 2), see section “Connecting several

pressure transducers”, pg. 37; further information upon request.

urther information upon request.

The configuration menu for specific settings, see section “Configuration”,

pg. 35, reflects automatically the connected components.

The types of components connected (e.g., valves) determine the ac-

tive menu items.

+

Maximum permissible pressure at the pressure trans-

ducer: 21.8 psi (1.5 bar) absolute.

VSK 3000:

+

Attention: At pressures above approxi-

mately 795 Torr (1060 mbar) the pressure

reading becomes incorrect due to satura-

tion of the pressure transducer. The display

flashes. Release pressure immediately!

Risk of bursting!

VSP 3000:
+

Attention: Maximum pressure reading:

7.5x10

2

Torr (1x10

3

mbar). Pressure values

above 7.5x10

2

Torr (1x10

3

mbar) are not dis-

played.

Risk of unnoticed overpressure.

Risk of bursting!