Automatic identification of connected components – VACUUBRAND DCP 3000 User Manual
Page 32
Automatic identification of connected components
When switching on the device, the configuration of the connected com-
ponents is checked.
Connected components (valves, pressure transducers, etc.) are au-
tomatically identified, used and monitored until the device is switched
off. Reconfiguration is possible by switching the DCP 3000 off/on. Identi-
cal components must be configured beforehand to different VACUU•BUS
addresses (e.g., VSK 1 and VSK 2), see section “Connecting several
pressure transducers”, pg. 37; further information upon request.
urther information upon request.
The configuration menu for specific settings, see section “Configuration”,
pg. 35, reflects automatically the connected components.
The types of components connected (e.g., valves) determine the ac-
tive menu items.
+
Maximum permissible pressure at the pressure trans-
ducer: 21.8 psi (1.5 bar) absolute.
VSK 3000:
+
Attention: At pressures above approxi-
mately 795 Torr (1060 mbar) the pressure
reading becomes incorrect due to satura-
tion of the pressure transducer. The display
flashes. Release pressure immediately!
Risk of bursting!
VSP 3000:
+
Attention: Maximum pressure reading:
7.5x10
2
Torr (1x10
3
mbar). Pressure values
above 7.5x10
2
Torr (1x10
3
mbar) are not dis-
played.
Risk of unnoticed overpressure.
Risk of bursting!