Cleaning the pressure transducer, Notice – VACUUBRAND DCP 3000 User Manual

Page 52

Advertising
background image

page 52 of 64

The vacuum gauge itself is maintenance free.
Condensate and deposits will affect the measurement results and/or the

adjustment of the pressure transducer.

Attention: Never use a pointed or sharp-edged tool to

clean the pressure transducer.

VSK 3000:

Never touch the ceramic diaphragm of the pressure trans-

ducer with hard objects.

VSP 3000:

The interior of the pressure transducer is highly sensitive!

Do not insert fingers or tools into the measuring chamber.

Fill the measurement chamber with a solvent (e.g., benzene) and al-

low sufficient cleaning time. Observe all regulations concerning usage

and disposal of solvents!

Drain the solvent and dispose of in accordance with regulations. Re-

peat cleaning if necessary.

Rinse the measurement chamber several times with alcohol in order

to remove all solvent residues.

Allow the pressure transducer to dry.

Readjust the pressure transducer if necessary, see sections “Read-

justment of VSK 3000”, pg. 54 / “Readjustment of VSP 3000”, pg.

56.

NOTICE

Cleaning the pressure transducer