4 installation examples, 1 measurement in an open pool or channel, Installation examples -6 3.4.1 – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual
Page 18: Measurement in an open pool or channel . 3-6
3 - 6
ba76040e04
05/2014
Commissioning
ViSolid
®
700 IQ (SW)
3.4
Installation examples
As a rule, the ViSolid
®
700 IQ (SW) will measure interference-free
when the distances and angles etc. specified are observed. However,
interferences at the measuring location (see section 3.3.2) may require
special adaptations of the installation.
3.4.1
Measurement in an open pool or channel
Measurement in a pool
The total suspended solids sensor can be suspended in the pool on a
chain (e.g. with EH/F 170 swing mounting assembly and EH/U 170
sensor holder). Make sure the sensor cannot bump against any walls
or obstacles.
Alternatively, the sensor can be immersed in the sample using a
pendulum mounting assembly, e.g. EH/P 170 pendulum mounting
assembly, (please note the minimum immersion depth).