Measuring
ViSolid
®
700 IQ (SW)
The calibration data entered are evaluated by the system. The entry of
the calibration value pairs can have the following results:
Results after entry of
Calibration data
17
Enter the value for the contents of total suspended solids (TSS
in mg/l) from the reference measurement with <> and
<OK>
.
18
Select the SiO2 value 1 menu item with <> and <OK>.
19
Enter the associated SiO2 value (SiO2 in mg/l) measured with
the ViSolid
®
700 IQ (SW) with <> and <OK>.
20
Repeat steps 13 - 18 until the required number of value pairs
(between 1 and 8) has been entered.
21
Terminate the entry of the calibration data with Save and quit.
To simplify the settings when entering the calibration data,
standardized high resolutions of 1 mg/l are possible. How-
ever, it is not sensible to use these in all cases.
Possible displays
Log book entries
(meaning/actions)
Measured value display
Successful new valid calibration value pairs
were entered for the sensor.
"----"
Calibration value pairs not accepted.
Sensor blocked for measurement.
– Enter the values again making sure that
they are entered in descending order
– Viewing the log book entry.
Information on the contents and structure of the log book
and how you can call it up is given in the L
OG
BOOK
chapter
of the IQ
S
ENSOR
N
ET
system operating manual.