YSI MultiLab 4010-3 User Manual
Page 104
Data storage
MultiLab 4010-3
104
ba76143e03
08/2013
Starting the
automatic storage
function
To start the automatic storage function, select continue with <><> and con-
firm with <ENTER>. The meter switches to the measured value display.
The active automatic data storage function can be recognized by the progress
bar in the status line. The progress bar indicates the remaining storage dura-
tion.
Terminating the
automatic storage
function prema-
turely
Proceed as follows to switch off the automatic data storage function before the
adjusted storage duration has expired:
1 Remaining storage duration
2 Graphical display of the storage duration
If the automatic data storage function is activated, only the follow-
ing keys are active: <M>, <STO_> and <On/Off>. The other keys
and the automatic switch-off function are deactivated.
01.08.2013 08:00
0d03h00min
1
01.08.2013 08:00
0d03h00min
2
1.
Press the <STO_> key.
The following window appears.
2.
Using <><>, select yes and confirm with <ENTER>.
The meter switches to the measured value display.
The automatic data storage function is terminated.
Stop automatic storage?
yes
no
Warning
01.08.2013 08:00