The default settings are – Starrett SR200 Surface Roughness Tester User Manual

Page 14

Advertising
background image

3.2

SR200

Specifications subject to change

Making a measurement

Note 1: If the pickup has been changed or the instrument is being used for the first time, the instrument

should be calibrated (see Chapter 7).

Note 2: Successful use of the SR200 will only be possible if it is operated on a surface free from external

vibration - see also operating notes in Chapter 5.

Switching the SR200 ON

Pressing the SCROLL  key brings the display on and the previously selected set-up is
displayed (provided power has been continuously present). The display is automatically turned
off if the instrument is not used for 30 seconds.

If the battery is dead or has been removed, the previously selected set-up is lost. When a
battery has been replaced and the SCROLL  key is pressed, the start up message will display
for 2 seconds and then the default settings are restored.

The default settings are

Parameter:

Ra

Cut-off:

0.8mm

Evaluation length:

4.0mm

Range:

100mm

Data dump evaluation length:

4.0mm

Data dump range:

100mm

No parameter/graph selected for printout
Language

English

Filter

Gaussian

If the user wishes to change any of these settings, this can be carried out using the SCROLL 
and SELECT  keys (see Chapter 4 for more detail)

Press the MEASURE  key. When the measurement is completed the pickup returns and the
results are displayed on the screen (see example below):

GaussLc

0.80mm

Eval Length 25.0mm
Ra

= 0.00µm

Rz

= 0.00µm

Rt

= 0.00µm

Rp

= 0.00µm

RSmm

= 0.00µm

more...>