Starrett SR200 Surface Roughness Tester User Manual

Page 23

Advertising
background image

SR200

Specifications subject to change

5.3

Measure before print pressing PRINT  key before measurement

No parameter selected pressing PRINT  key before selecting printout parameters
Printer not connected

pressing PRINT  key when printer not connected or no computer
connected to receive dump data.

Printing cancelled

pressing PRINT  key during data transmission to printer (stop
printing). Message displayed for 2 seconds.

Measurement
cancelled

Pressing MEASURE  key during traverse (stop traverse)

Measure before data
dump

Attempted data dump before measurement

Specification

Battery

Alkaline: minimum 600 measurements of 4mm measurement length

Ni-Cad: minimum 200 measurements of 4mm measurement length

Size: 6 LR 61 (USA/Japan)

Traverse Speed

1mm/sec

Measurement

metric/inch units

Cut-off values

0.25mm, 0.8mm and 2.5mm (0.01in, 0.03in and 0.1in)

Traverse lengths

0.25-25mm (0.05-0.98in)

Display

LCD-matrix, 8 lines x 20 characters, alphanumeric

9 pin D-connector female. Containing RS232 hardwired handshake
(RS232 level).

Accessory socket

Transmit function only hardware implemented.

Remote start (5V logic)

Remote start: short to ground.

Constant power on instrument (5V logic)

When connected to Ground, constant power is on

RS232 Output

Printer Configuration

If SPC is on

Baud rate

4800 Baud

Parity

Even

Data bit length

7 bits

Printer head disabled

If SPC is off

Baud rate

9600 Baud

Parity

Odd

Data bit length

8 bits

Printer head enabled

Output to printer (in ASCII characters) or pre-selected parameters and provided that G is pre-

selected, a readout of dc corrected surface profile.

Advertising