Section 4

OPERATION

Performing Ionizer Tests

Initiate automated test sequences and manual tests directly from the system prompt,

-.-.

Selecting a Manual Test:

Press

+Decay to start a positive discharge test.

Press

-Decay to start a negative discharge test.

Press

Bal to start a balance (offset voltage) test.

The respective indicator is on and flashing while a test is in
progress.

The respective indicator is on and not flashing when the test is
complete and data is available.

Manual Zero for the Plate:

In the event that the instrument does not indicate zero volts for
a balance test (when the ion collecting plate is hard-grounded),
the fieldmeter can be zeroed by pressing the

Bal button for >2

seconds. This would be done from the system prompt,

-.-.

Selecting an Automated

Test Sequence:

Press

Auto to run a sequence of discharge and balance tests.

Cancelling a Test:

Press

Shift/Cancel to cancel a test after it is started.

To View Test Results:

Test results are displayed after each test.

For a Decay test, the Decay time is displayed until a new
button is pressed. In the instance when a decay test fails, such
as if there is no ionization to reduce the plate voltage, a failure
message is displayed: either

+DCY DFAI or -DCY DFAI,

depending on the polarity of the decay test.

For a Balance test, the display scrolls through the negative
peak voltage (

VMIN), the positive peak voltage (VMAX) and the

average voltage (

AVG).

Press

+Decay, -Decay, Bal or Auto to start a new test or

sequence of tests.

Press

Shift/Cancel to return to the system prompt -.-.

Test data can also be viewed using the

ReCall feature of the

Model 300.