Transcutaneous Pacemaker Testing
Setup
and
Testing
3
3-27
F1
F2
F3
F4
eyr025.eps
2. Press
F1 OUTPUT
to toggle among four simulated operational modes:
•
Demand
•
Continuous
•
Non-capture
•
Non-function.
3. Press
F3 THRESH
UP
or
DOWN
arrow key to adjust the threshold
amplitude.
4. Press
F4 NSR RATE UP
or
DOWN
arrow key to adjust the NSR rate.
Note
The NSR rate function is active only during the demand mode of
operation.
Long Term Tests
The overall stability of the pacemaker’s pulse rate and amplitude are
monitored continuously over an extended period of time. The user enters the
desired target values for these two parameters into the Analyzer. This program