Transcutaneous Pacemaker Testing
Setup
and
Testing
3
3-29
3. For each, pulse rate and amplitude, press
F3 LIMIT
UP
or
DOWN
arrow
key to select the limit value.
4. Press
F4
LOAD UP
or
DOWN
arrow key to select the load.
5. Press
F1
NEXT
to advance the screen.
F1
F2
F3
F4
eyr027.eps
6. Press
F3
AMP
UP
or
DOWN
arrow key to adjust the amplitude.
7. Press
F4
RATE
UP
or
DOWN
arrow key to adjust the rate.
Note
Target values for amplitude and rate must be entered for the test to
perform correctly.
8. Press
F1
START
to advance the screen and start the test.