Troubleshooting using gauge readings, Re reading is 25-28 psi pressure. see, Figure 5-3 on – Gasboy Gilbarco Global Pumping Unit Operation Manual User Manual

Page 28

Advertising
background image

Troubleshooting

Troubleshooting Using Gauge Readings

Page 24 MDE-4447A Global Pumping Unit Operation and Service Manual · May 2008

Figure 5-3: Pressure Reading - Nozzle Closed

Troubleshooting Using Gauge Readings

The following examples help you determine the possible cause of a problem by observing the
gauge readings on the inlet (vacuum) and outlet (pressure) side of the pumping unit. Actual
readings may vary slightly, depending on installation and environmental conditions.

Figure 5-4: Gauge Troubleshooting 1

Symptoms: No Flow; Nozzle Open

Check Inlet Side

Probable Causes:
• Control Valve Stuck Shut
• Bypass Valve Open
• Atmospheric Float Valve Open
• Stuck Rotor Blades
• Broken Suction Line
• Strainer or Filter Completely

Plugged or in Backwards

• Empty Tank

At Control Valve
Cover

At Strainer
(or Filter)
Cover

Advertising