K-Patents PR-03 User Manual
Page 93
93
93
93
93
8 Sensor specifications and other sensor information
87
8.7.2 PR-03-W specifications
General specifications
General system specifications for a PR-03-W refractometer sensor
Refractive index range, standard
R.I. 1.3100–1.5400 (corresponds to hot water–100 Conc% b.w.) with
spinel prism
Refractive index range, option
R.I. 1.2600–1.5000 with sapphire prism
Accuracy
R.I. ± 0.0002 (corresponds typically to ± 0.1% by weight).
Repeatability and stability correspond to accuracy.
Speed of response
1.2 s undamped
Damping time constant
Selectable up to 5 min
Process temperature
-20
◦
C–100
◦
C (-4
◦
F–212
◦
F)
Temperature compensation
Automatic
Ambient temperature
max. 45
◦
C (113
◦
F), min. -20
◦
C (-4
◦
F)
Humidity
20–100 %
Pollution degree
Degree 2
Process pressure
max. 10 bar (145 psi)
Recommended flow velocity
1.5–6 m/s
Protected by
US Patent No. US6067151, German Patent No. DE19855218
Sensor
Wetted parts, standard
Teflon
R
PTFE, prism gasket Teflon
R
PTFE, prism spinel or sapphire,
O-ring Kalrez, adaptor sapphire
Sensor weight
With 2” Saunders body 15 kg (33 lbs)/3” Saunders body 26 kg
(57 lbs)/4” Saunders body 33 kg (73 lbs)
Sensor protection class
IP67, Nema 4X
Process connection
With PFA (Fluorinated ethylene propylene) lined Saunders valve body
2", 3" or 4"
Temperature measurement
Built-in Pt-100
Image detector
1024 pixel CCD-element
Light source
Light emitting diode (LED)
Sensor housing
AISI 316L stainless steel cover
Exproof
Exproof by purging
Saunders valve body
Valve body material
Graphite cast iron
Valve body lining material
PFA (Fluorinated ethylene propylene)
Process connection
ANSI flange 4 inch 150 lbs / ANSI flange 3 inch 150 lbs /
ANSI flange 2 inch 150 lbs / DIN flange DN 100 PN 16 /
DIN flange DN 80 PN 16 / DIN flange DN 50 PN 16
Indicating transmitter
See Chapter 9
Options
− External output unit, Section 4.2
− Relay units for alarm, Section 4.1