Air-Oxygen Blender

USA and CANADA ONLY

LOW Flow Operation Verification Procedure

(50 psi / 3.45 bar MODELS)

Serial No.

S
E

Q
#

DIAL

SET

Oxygen%

OXYGEN

PRESS

±1.0

AIR

PRESS

±1.0

FLOWMETER

SET TO

lpm ±0.2

AUXILIARY

BLEED

FUNCTION

TARGET

VALUE

ACTUAL

VALUE

psi

bar

psi

bar

1

ANY

50

3.45

50

3.45

closed

closed

leak

<2 psi

/

2 MIN

*2

60

75

5.17

0

0

0

closed

back flow <100 ml/min

*3

60

0

0

75

5.17

0

closed

back flow <100 ml/min

4

21

50

3.45

50

3.45

3

open

end point

(±0.3)

Source Value

5

40

50

3.45

50

3.45

3

open

set point 37.0%-43.0%

6

60

50

3.45

50

3.45

3

open

set point 57.0%-63.0%

7

80

50

3.45

50

3.45

3

open

set point 77.0%-83.0%

8

100

50

3.45

50

3.45

3

open

end point

(±0.3)

Source Value

9

60

50

3.45

50

3.45

1

open

set point 57.0%-63.0%

10

60

60

4.14

67

4.62

1

open

set point 57.0%-63.0%

11

60

60

4.14

50

3.45

1

open

set point 57.0%-63.0%

12

60

50

3.45

Slowly

reduce

to 30

Slowly

reduce

to 2.07

3

closed

Alarm

ON

30.0

± 2.0 psi

13

60

50

3.45

Slowly

Increase

until alarm

shuts off

3

closed

Alarm

OFF

45.0 psi

MAX

14

60

Slowly

Reduce

to 30

Slowly

Reduce

to 2.07

50

3.45

3

closed

Alarm

ON

30.0

± 2.0 psi

15

60

Slowly

Increase

until alarm

shuts off

50

3.45

3

closed

Alarm

OFF

45.0 psi

MAX

16

60

50

3.45

50

3.45

MAX

closed

flow rate

30.0 lpm

MIN

17

60

50

3.45

0

MAX

closed

flow rate

30.0 lpm

MIN

18

60

0

50

3.45

MAX

closed

flow rate

30.0 lpm

MIN

19

60

50

3.45

50

3.45

MAX

open

flow rate

30.0 lpm

MIN

* Reference, Letter D. (Reverse Gas Flow Procedure)