High flow – Precision Medical Air-Oxygen Blender User Manual

Page 19

Advertising
background image

- 18 -

Air-Oxygen Blender

HIGH FLOW

Figure C

8. Remove dial [1] by pulling dial away from

manifold block [8].

9. Remove the auxiliary [10] outlets (auxiliary

contains Blue Muffler [38]) by using a ½

in. open end wrench to unscrew.

10. Use Retaining Ring Pliers to unscrew and

remove audio alarm assembly [11] from the

bottom of the manifold block.

[8]

[10]

[11]

[1]

[38]

contained in [10]

Figure D

11. Use small retaining ring pliers to remove

retaining rings [12] from each side of

the alarm assembly [13]. Push the alarm

assembly through to remove the assembly

from manifold block.

[12]

QUANTITY - 2

[13]

Figure E

12. Using a 11/32 in. nut driver or socket, loosen

the nut [14] (ONLY two turns) holding the

knob guide [15] on the proportioning valve

assembly [16]. Slide knob guide assembly

from proportioning valve assembly shaft.

Slide resistance ring [17] from proportioning

valve assembly.

13. Remove phillips head screw [18] and

washer [19] from side of proportioning valve

assembly.

14. Replace knob guide assembly and tighten

the nut.

15. Pull knob guide assembly to remove

proportioning valve assembly from manifold

block. Again, loosen the nut (ONLY two turns)

holding the knob guide to the proportioning

valve assembly, remove knob guide from

proportioning valve assembly shaft.

[14]

[15]

[16]

[18]

[19]

[17]