Portable Oxygen Concentrator

CAUTION

• Inspect the POC for visual damage before use, DO NOT USE if damaged.

• To

prevent damage to the POC, DO NOT operate the POC without

the inlet filter or while the inlet filter is wet.

Using the EasyPulse POC

1. Verify battery status (level) prior to each use, charge as necessary.

2. Connect the end of a standard adult single lumen oxygen nasal cannula

with a maximum length of 7 feet (2.1 m) to the POC’s oxygen outlet fitting.

To ensure proper device operation, use only

an adult single lumen cannula maximum 7

ft (2.1 m).

3. Place the cannula over your ears and position

the prongs in your nose as instructed by your

Equipment Provider or cannula manufacturer.

OR

4. Prior to operation, ensure the inlet and exhaust vents on the POC are clear.

Any blockage of the vents can inhibit performance.

5. Choose one of the 3 power options: AC Power Adapter, DC Power Adapter or

Internal Battery (Refer to the “POWER SUPPLY” section of the user manual

for details on these 3 options).

6. Turn “ON” the POC by rotating the setting knob clockwise until the knob points

to your prescribed flow setting number and the number is illuminated green.

7. Breathe normally through your nose. A measured pulse of oxygen will be

delivered each time the POC senses an inhalation. To show the POC is

triggering properly the illuminated setting number will turn off briefly each

time it senses an inhalation.

8. To turn “OFF” the POC, rotate the setting knob counterclockwise to the

“OFF” ( ) position.

WARNING

ISO 80601-2-69 (Standard for Oxygen Concentrators) highly

recommends that the patient cannula that delivers gas to the patient

from the oxygen concentrator should include a Fire Stop Check Valve

to stop the flow of gas towards the patient in the case that the cannula

becomes ignited. The Fire Stop Check Valve should be located as

close to the patient as reasonably practicable.

Fire Stop Check Valve - Precision Medical Part # 507706

2” Tubing Connector - Precision Medical Part # 507707