Component description, Operating instructions, Caution – Precision Medical PM4130 EasyPulse POC3 User Manual
Page 9: Warning
8
Portable Oxygen Concentrator
CAUTION
Missing or illegible labels must be replaced, contact Precision Medical, Inc.
COMPONENT DESCRIPTION
Product
Serial # Label
Inlet Filter
Setting
Indicator
Power Input
Jack
Handle
Setting Knob
Oxygen Outlet
Fitting
Internal Battery
Level Indicator
Alarm Indicator
OPERATING INSTRUCTIONS
WARNING
• Read this User Manual before operating the POC.
• The POC shall be used by an adult or supervised by an adult when
being used by a child.
• Additional monitoring or attention may be required for patients using this
POC who are unable to hear or see alarms or communicate discomfort.
• The Setting Knob does not rotate 360°. Rotating the knob less than
the “OFF ( ) position” or over “setting 3” will damage the Concentrator.
CAUTION
•
ALWAYS confirm prescribed dose before administering to patient
and monitor on a frequent basis.
• For Proper Oxygen Flow, be sure that cannula is not kinked or
obstructed before or during use.
•
DO NOT use humidifier bottles with the POC.
•
DO NOT use pediatric, low flow nasal cannulas or oxygen masks
with the POC.