WARNING
• ALWAYS confirm prescribed dose before administering to
patient and monitor on a frequent basis.
• Always follow ANSI and CGA standards for Medical Gas
Products and Oxygen Handling (G-4).
• Keep cylinder valve closed at all times when cylinder is not in use.
• NO OXYGEN is delivered when the pointer is aligned with
“OFF”, or between settings.
• This Conserver is not to be used by patients who breathe
through their mouths.
• DO NOT use if dirt or contaminants are present on or around
cylinder, valve, product or connecting devices.
• DO NOT use oils, greases, organic lubricants or any
combustible materials on or near this product. Wash and dry
hands properly prior to usage.
• DO NOT use a humidifier with the Conserver.
• DO NOT allow cylinders to tip or fall. Secure gas cylinders
so they cannot fall. For optimum safety keep cylinder upright
whenever possible.
• DO NOT store cylinders near sources of heat or flame.
• DO NOT use while sleeping without consulting your Provider.
• DO NOT smoke in an area where oxygen is being administered.
• DO NOT use near any type of flame or flammable/explosive
substances.
• Use only USP grade medical oxygen.
• The Conserver is equipped with a relief valve. If you hear
a loud hissing or popping sound coming from the product,
discontinue use of product, close cylinder valve, and contact
your Provider.
• The Conserver is designed to operate with a single lumen,
adult cannula with a maximum length of 7 feet (2.1 m).
• The cannula is for single patient use only.