Energizer 715 User Manual
Page 115

Optimizing pump therapy 93
1
From the HOME screen, press
. The SET EASY BOLUS
screen will appear. The single “step” value will be flashing.
2
Press
the number of times needed for your bolus
amount. Watch the amount change on the screen with each
press. When your total bolus amount appears on the screen,
press
ACT
. The pump will vibrate or sound a different tone
for each
press. Listen/feel to count the steps without
looking at the screen.
For example: You need to deliver a 1.0 unit bolus, and your step size is set to 0.1. Each time you
press
, the units increase by the “step” amount. To deliver 1.0 units, you need
to press the
button 10 times. (10 X 0.1 = 1.0) The screen will show 1.0 units.
NOTE -
Pressing or
ESC
will cancel the Easy bolus. For your safety you cannot use the
to select the Easy Bolus values.
3
If this amount is correct, press
ACT
to start the
Easy Bolus delivery. The BOLUS DELIVERY screen
will show the units being delivered. When the
total bolus is finished, the pump will beep or
vibrate.
If this amount is wrong, press
ESC
or
to
start over. The pump will return to the HOME
screen.
(flashing)
SET EASY BOLUS
U
0.1
x 10
SET EASY BOLUS
U
ACT
1.0
SET EASY BOLUS
1.0
U
0.1
U
BOLUS DELIVERY
1:32P
Normal
ACT
to accept
ESC
to cancel