Selecting temp basal type

Your pump will remember the temp basal type setting. Once the type is set, you do not have to set

it again. To select a temp basal type, do these steps:

1

Go to the BASAL MENU.

Main > Basal > Temp Basal Type

Select

Temp Basal Type

and press

ACT

.

2

The SET TEMP BASAL AS screen will appear. Select

Insulin Rate

or

Percent of

Basal

and press

ACT

.

3

The screen will return to the BASAL MENU. The temp basal type is now set. Exit
the menus.

NOTE -

Be aware. If your temp basal type is set to “Percent of Basal,”

changes to your basal rate are not allowed until after temp basal is
completed or cancelled.

Delivering a temp basal

NOTE -

A temp basal cannot exceed your programmed max basal rate.

1

Go to the BASAL MENU.

Main > Basal > Set/Edit Temp Basal

Select

Set/Edit Temp Basal

and press

ACT

.

2

The SET DURATION screen will appear. The duration* will flash. Enter the
desired minutes or hours (30 minutes to 24-hours), then press

ACT

.

NOTE -

* Duration is the amount of time it will take for the pump to deliver

the temporary basal.

ACT

BASAL MENU

Max Basal Rate

Patterns

Temp Basal Type

SET TEMP BASAL AS

ACT

Insulin Rate <U/H>

Percent of Basal

BASAL MENU

Max Basal

Rate Patterns

ESC

Temp Basal Type

BASAL MENU

Set/Edit Temp

Basal Set/Edit
Basal Review

ACT

SET DURATION

3:00

Temp

ACT