Display "custody mode" (0a9), Active pos, Active neg – Lightning Audio 7532 User Manual

Page 59

Advertising
background image

59

7532

Commissioning

Display "custody mode" (0A9)

This indicates the instrument calibration mode. The calibration mode (active) can be set

using the hardware security lock on the electronics (see Page 34).

Selection:
• inactive
• active pos.
• active neg.

active pos.

The custody mode (instrument is lead-sealed and accurate to the nearest mm) is active

and is held.

active neg.

Custody mode (instrument is lead-sealed and accurate to the nearest mm) is activated

and not held, e.g. because the signal-to-noise ratio is less than 10 dB (refer to

function "echo quality" (056) in the function group "extended calibr." (05)).

Caution!

After entering all the values and completing mounting and aligning work, enter the

Reset Code "555" in the function

"history reset" (009)

of the "basic setup" (00) function

group to reset the instrument history for auto-correction.