278 VT1422A Command Reference

Chapter 7

DIAGnostic

DIAGnostic:CALibration:TARE[:OTDetect]:MODE

DIAGnostic:CALibration:TARE[:OTDetect]:MODE <mode>

sets whether Open

Transducer Detect current will be turned off or left on (the default mode) during the
CAL:TARE operation.

Parameters

Comments

When <mode> is set to 0 (the *RST Default), channels are tare calibrated with

their OTD current off. When <mode> is 1, channels that have their OTD current
on (DIAGnostic:OTDetect ON,(@<ch_list>)) are tare calibrated with their
OTD current left on.

By default (*RST) the CALibration:TARE? command will calibrate all

channels with the OTD circuitry disabled. This is done for two reasons.
First, most users do not leave OTD enabled while taking readings; and second,
the CALibration:TARE? operation takes much longer with OTD enabled.
However, for users who intend to take readings with OTD enabled, setting
DIAG:CAL:TARE:OTD:MODE to 1, will force the CAL:TARE? command to
perform calibration with OTD enabled on channels so specified by the user with
the DIAG:OTD command.

Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?

*RST Condition: DIAG:CAL:TARE:MODE 0

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

configure OTD on during CAL:TARE

DIAG:CAL:TARE:MODE 1

set mode for OTD to stay on

CAL:TARE?

start channel tare cal.

DIAGnostic:CALibration:TARE[:OTDetect]:MODE?

DIAGnostic:CALibration:TARE[:OTDetect]:MODE?

returns the currently set

mode for controlling Open Transducer Detect current while performing CAL:TARE?
operation.

Comments

Returns a 0 when OTD current will be turned off during CAL:TARE?. Returns

1 when OTD current will be left on during CAL:TARE? operation. The data
type is int16.

Related Commands: DIAG:CAL:TARE:MOD, DIAG:OTD, CAL:TARE?

*RST Condition: DIAG:CAL:TARE:MODE 0

Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

mode

boolean (uint 16)

0 | 1

volts