Error Messages 457

Chapter B

3053

'Corrupt on board Flash memory.'

3056

'Custom EU not loaded.' May have erased custom
EU conversion table with *RST. May have linked
channel with standard EU after loading custom EU, this
erases the custom EU for this channel. Reload custom
EU table using DIAG:CUST:LIN or DIAG:CUST:PIEC.

3057

'Invalid ARM or TRIG source when S/H SCPs
enabled.' Don't set TRIG:SOUR or ARM:SOUR to
SCP with VT1510A or VT1511A installed.

3058

'Hardware does not have D32, S/H or new trigger
capabilities.' Module's serial number is earlier than
3313A00530.

3067

'Multiple attempts to erase Flash Memory failed.'

3068

'Multiple attempts to program Flash Memory failed.'

3069

'Programming voltage jumper not set properly.'
See "Disabling Flash Memory Access" in Chapter 1
(JM2201).

3070

'Identification of Flash ROM incorrect.'

3071

'Checksum error on Flash Memory.'

3074

'WARNING! Old Opt 16 or Opt 17 card can damage
SCP modules' must use VT1506A or VT1507A.

3075

'Too many entries in CVT list.'

3076

'Invalid entry in CVT list.' Can only be 10 to 511.

3077

'Too many updates in queue. Must send UPDATE
command.' To allow more updates per ALG:UPD,
increase ALG:UPD:WINDOW

3078

'Invalid Algorithm name.' Can only be 'ALG1' through
'ALG32' or 'GLOBALS' or 'MAIN'.

3079

'Algorithm is undefined.' In ALG:SCAL, ALG:SCAL?,
ALG:ARR or ALG:ARR?

3080

'Algorithm already defined.' Trying to repeat ALG:DEF
with same <alg_name> (and is not enabled to swap) or
trying to define 'GLOBALS' again since last *RST.

3081

'Variable is undefined.' Algorithm exists but has no
local variable by that name.

3082

'Invalid Variable name.' Must be valid 'C' identifier.
See Chapter 7.