Programming the VT1422A & VT1529A/B for Remote Strain Measurement 93

Chapter 3

Error Analysis

Figure 3-21 compares the non-linear quarter bridge equation used for strain
SCPs with the linear approximation used with the VT1529A/B. Notice that
while the error is independent of excitation voltage and unstrained voltage,
error is quite sensitive to gage factor.

Figure 3-21. Error of Quarter Bridge Linear Approximation

Error vs Delta Strain

-1000

0

1000

2000

3000

4000

5000

6000

-100000

-80000

-60000

-40000

-20000

0

20000

40000

60000

80000

100000

Delta Strain in m icrostrain (µ ε)

Error in microstrain (µ

ε)

GF=1
GF=2
GF=5