INSTALLATION AND POWER-UP

2-4

1

RF OUT

50

Ω SMA-type socket. Protected against the application of reverse power of up to

50 W.

2

SYSTEM FAIL

This red LED lights to indicate that the signal generator has failed, or is in the process

of executing its self-test. It indicates the condition of the VXI-bus SYSFAIL line.

The LED will continue to be lit whilst the self-test is in progress even when SYSFAIL

is inhibited by the commander.

3

POWER OK

This green LED lights to indicate that power is being supplied to the signal generator.

All lines are continuously checked for sufficient voltage.

4

RPP TRIP

This red LED lights to indicate that the Reverse Power Protection (RPP) circuit has

tripped. The power source must be removed from the RF OUT socket.

5

FREQ STD I/O

BNC socket for the input of external standard frequencies of either 1 MHz or

10 MHz. Also supplies a 10 MHz internal standard output.

6

EXT MOD INPUT

BNC socket which allows an external modulating signal to be applied.

7

TRIGGER INPUT

BNC socket which has three uses; in priority order these are:

FSK logic input

Memory

sequencing

Sweep

trigger.

8

PULSE INPUT

10 k

Ω BNC socket which accepts a pulsed input. Also used as one logic input (the

other is the TRIGGER INPUT) for 4FSK modulation.

9

LF OUTPUT

600

Ω BNC socket which monitors the modulation oscillator.

Switching on

Insert the signal generator module in the required slot in the mainframe and screw in the retaining
screws. Switch the mainframe on. All three LEDs should initially light while the generator carries
out its self checks. When the unit passes its self checks, and if there are no errors detected on the
backplane, the red SYSTEM FAIL and RPP LEDs will go out within 5 seconds and the green
POWER OK LED will remain on.

Disk installation/loading instructions

LabWindows/CVI Instrument Driver and VXI Plug and Play Soft Panel disks are supplied with this
instrument. Before inserting a disk in your disk drive read the installation or loading instructions
given on the label of the appropriate disk. Refer to 'read me' files for further information.