5-7

Spectral purity

Specification

Harmonics

Typically better than

−30 dBc for levels up to +7 dBm,

typically better than

−25 dBc for levels 6 dB below the maximum specified output.

Non-harmonics (offsets >
3 kHz)

Better than

−70 dBc for carrier frequencies up to 1 GHz,

better than

−64 dBc for carrier frequencies above 1 GHz,

better than

−60 dBc for carrier frequencies above 2 GHz.

Residual FM (FM off)

Less than 4.5 Hz RMS in a 300 Hz to 3.4 kHz bandwidth at a carrier frequency of
1 GHz.

SSB phase noise

Better than

−124 dBc/Hz at 20 kHz offset from a 470 MHz carrier.

Typically

−121 dBc/Hz at 20 kHz offset from a 1 GHz carrier.

Test equipment

Description Minimum

specification Example

Spectrum
analyzer

DC to 7.2 GHz frequency coverage

Anritsu MS2602A

Measuring
receiver

0 dBm to

−127 dBm; 2.5 MHz to 2.4 GHz Capable

of measuring residual FM less than 2 Hz and SSB
phase noise <

−124 dBc/Hz at 20 kHz offset from a

1 GHz carrier

Agilent 8902A
with option 037

Signal generator

+8 dBm from 32.5 MHz to 2.43 GHz

IFR 2041

Harmonics

Test procedure

C5357

RF OUT

RF INPUT

UUT

Spectrum analyzer

Fig. 5-4 Carrier harmonics and non-harmonics test set-up

(1) Press CAL on the spectrum analyzer.

(2) Connect the test equipment as shown in Fig. 5-4.

(3) On the UUT set:

Carr Freq

10 kHz

RF

Level

−4 dBm