Routine tasks, After running samples, Sheath and waste fluids – Luminex 200 User Manual with IS 2.3 Software User Manual

Page 29

Advertising
background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-144 Rev. B

3 - 3

5. Loosen the probe adjustment screw on the probe holder one-third

to one-half turn. Pull it upward until it touches the top of the
adjustment slide. Tighten the thumbscrew.

6. Click Sample Probe Down to lower the sample probe.

7. Gently push the probe down until it just touches the top of the

alignment discs or sphere.

8. Tighten the probe adjustment screw.

9. Click Sample Probe Up to raise the sample probe.

10. Replace the plastic shield that covers the sample probe area.

After Running Samples

‹

After running samples:

1. Sanitize with a 10% to 20% household bleach solution.

2. Run two Wash cycles with distilled water.

3. Soak with distilled water. Wait until the soak completes.

4. If desired, turn off the Luminex 200 analyzer.

Routine Tasks

Sheath and Waste Fluids

Replace the sheath fluid and empty the waste container as required.
Be careful not to touch the seal in the waste bottle cap or allow it to
become wet or dirty. This could cause the bottle to become
pressurized, causing pressure errors in the system. If the seal gets
wet, let it air-dry. Touching the seal can contaminate it. If it gets
dirty, you should replace the waste bottle cap.

If you are using the waste line, do not move it while the system is
running. While you can move the line along a horizontal surface, do
not permanently reroute the waste line elevation without first
contacting Technical Support. You can move the line temporarily for
cleaning and maintenance purposes.

You must manually monitor waste container levels. The Run Batch
tab displays a warning when you need to refill (or replace) the sheath
fluid container. See Figure 3-4.