Fluid Leaks

Fluid leaks can result in poor pressurization and failed sample
acquisition.

Sample Probe

Problems with the sample probe can lead to fluid leaks and
pressurization problems, as well as inhibit sample acquisition.

Symptom

Possible problem

Solution

Pressure too low.

The sample probe is clogged.

Clean the sample probe. See
page 3-5.

The syringe seal leaks.

Replace the syringe seal. See
page 3-9.

The syringe valve leaks.

Hand-tighten the syringe
connection (silver knob) on the
syringe valve. Run a Prime
command. If leaks continue,
contact Technical Support.

Large amount of fluid pooled
around instrument.

Fittings or fluid lines are
damaged.

Contact Technical Support.

Fluid dripping from the
sample probe.

The sample probe is clogged.

Clean the sample probe. See
page 3-5.

The sample three-way valve is
faulty.

Contact Technical Support.

Fluid is leaking from the front
of the analyzer.

The syringe seal leaks.

Replace the syringe seal. See
page 3-9.

The syringe valve leaks.

Hand-tighten the syringe
connection (silver knob) on the
syringe valve. Run a Prime. If
leaks continue, contact
Technical Support.

Symptom

Possible problem

Solution

Sample probe leak.

The sample probe is clogged.

Clean the sample probe. See
page 3-5.

Sample arm is stuck in the up
position

The system isn’t properly
pressurized.

Check pressure settings. Ensure
that the sample probe is not
clogged and there are no leaks
in the syringe seal or syringe
valve.