10.2

Cell HolDer MoDUleS

A number of cell holder module options are available for use with these instruments:

• 10x10mm cell holder
• 8 cell turret
• Water heated turret
• Test tube holder

These are supplied as a complete plug-in module as detailed above. To fit these modules the following
procedure should be used:

Disconnect the instrument from the mains supply.

Check the installed sample chamber module and remove any filled cuvettes to avoid spillage.

Raise the lid and loosen the sample chamber retaining screw located at the front of the module. Lifting
the front carefully remove the module from the instrument and store safely for possible future use.

Unpack the new module and check to ensure no damage has occurred during transit.

To install the new module raise the lid by 30° and gently ease it into the instrument compartment
taking care that the lugs shown in the diagram locate into the main casework.

Once the module is in position, tighten the sample chamber retaining screw.

Re-connect the mains supply to the rear panel of the instrument. The instrument is now ready for use
with the new module.

127

Raise lid by 30°

for re-assembly

Ensure lugs locate

into main case

Sample chamber

retaining screw