131

If the module has been successfully installed the instrument will show

‘Passed’ alongside the

Accessory test… indicator.

If the module is incorrectly fitted to the instrument the message

‘none Found’ will be displayed. If

this occurs switch the instrument off and disconnect from the mains supply. Remove the module and
carefully refit. Reconnect the instrument to the mains supply and switch on. If the fault remains please
contact your local dealer for advice.

If the carousel is incorrectly fitted the message

‘Unable to calibrate multicell holder’ will be

displayed and the

Accessory test… will show ‘Failed 08’. In this instance switch the instrument off

and fit the carousel, ensuring correct alignment as detailed below.

If the

‘Failed 08’ message is displayed again, switch the instrument off and disconnect from the mains

supply. Remove the module and check for any obvious signs of damage. Carefully refit the module
following the fitting procedure and power up the instrument. If the instrument fails again please
consult your dealer for advice.

Manual operation

The following operating parameters apply to all modes with the exception of Kinetics.

When the 8 or 6 cell changer is fitted two additional icons will be displayed. The icon at the top of the
screen in the status bar indicates the current cell position.

The icon at the bottom of the display allows the user access to an input screen for selection of cuvette
positions.

Automatic operation

Selecting the

Accessory icon from the Method Set-up options opens the tabbed page for the

accessory settings.

The setting options are identical for the Photometrics, Spectrum and Multi-Wavelength modes. The
Quantitation mode is similar with just the addition of a manual or automatic option for the generation
of calibration curves. Kinetics is fully automated for multiple, parallel kinetic assays and simply requires
the input of the number of samples to be run simultaneously.

Automatic Multi-Cell

Selecting the

Automatic Multi-Cell key alternates the setting between Enabled and Disabled. In the

latter case all the other options are irrelevant and are greyed out, the automatic multi-cell changer can
then only be used in the

Manual mode as described above. With Enabled selected the other options

can be set to the desired status for the current method.