Storage of explosive and
highly fl ammable substances

Laboratory refrigerators and freezers with Comfort electronic controller and spark-free interior

T

he range of laboratory appliances with Comfort electronic

controllers comprises two refrigerators and two freezers with

spark-free interior for storing explosive and highly fl ammable

substances - for instance in the chemical industry or in special

laboratories. The interiors of the appliances are in compliance with

the safety requirements of the EU Directive 94/9/EC (ATEX 95) and

have been tested to the EN 1127-1, IEC 60079-0 and IEC 60079-15

standards by electro-suisse – SEV (Swiss Association for Electrical

Engineering, Power and Information Technologies), the ATEX

conformity evaluation organisation.

26-27_LaborSpecial_EN.indd 26

11.02.14 12:36

The Comfort electronic controller

The precision Comfort electronic controller with digital temperature display enables temperatures to be

set accurately. Symbols indicate the operating status of the appliance at all times. To satisfy laboratory

hygiene requirements, the Comfort electronic controller is fl ush-mounted and has an easy-clean

membrane keypad.

Certifi ed to ATEX 95

All laboratory refrigerators and freezers with spark-free interior are tested according to the 94/9 EC

(ATEX 95) Directive. Rated as II 3G Ex nA II T6, these appliances are suitable for storing explosive and

highly fl ammable substances in sealed containers.

Maximum temperature stability and temperature consistency

In conjunction with the precision Comfort electronic controller, the forced-air cooling system of the

laboratory refrigerators and the static cooling system of the laboratory freezers with spark-free

interior ensure maximum temperature stability and high temperature consistency. The laboratory

refrigerators are equipped with a secondary safety thermostat to prevent temperatures below +2°C in

case of a fault. All the laboratory appliances with Comfort electronic controllers are tested according

to the NF X 15-140 standard with regard to maximum temperature stability and optimal temperature

consistency.

Integrated alarm systems

A visual and audible temperature alarm alerts the user if temperature deviation limits are

exceeded. The parameters for the temperature alarm can be individually set. In addition, a visual

and audible door alarm is activated if the door is open for longer than 1 minute. The door-open alarm

delay can be adjusted between 1 and 5 minutes. Furthermore, the laboratory appliances with

Comfort electronic controllers have a visual power failure alarm when mains power returns and

a sensor defect alarm.

Integrated data memory with min/max temperatures

The Comfort electronic controller has an integrated data memory. It logs the maximum and minimum

internal temperatures that occurred once the set temperature was reached, the last three temperature

alarms and power failures, together with the time, date and duration of the alarm. The respective data

can be recalled using the AlarmLog function and read on the display panel.

1-point calibration

The laboratory appliances with Comfort electronic controllers provide 1-point calibration for accurate

temperature control. This permits compensation between the set temperature and the actual internal

temperature. Positive or negative changes to the correction value can be made in steps of 0.1K.