Laboratory refrigerators and freezers with Profi electronic controller

Energy-effi cient hot-gas defrost system

The hot-gas defrost system with demand and time control permits very short defrost cycles: only

8 minutes for laboratory refrigerators and 12 minutes for laboratory freezers. As a result, the internal

temperature remains almost constant during the defrost cycle. In addition, there is the option of initiating

the defrost manually.

External temperature and alarm documentation

The laboratory appliances with Profi electronic controllers are equipped with a volt-free contact for

alarm forwarding to an external remote warning system. The data stored in the internal memory can be

read out via the infrared interface and transferred to a PC with documentation software. In addition, the

appliances have an RS 485 serial interface enabling as many as 20 appliances to be networked with a

central documentation and alarm system.

High-quality interior for versatile use

The smooth compartment liners are made entirely of 304-grade stainless steel and are therefore

particularly easy to clean. The robust trayslides can be easily repositioned to suit needs and, together

with the plastic-coated grid shelves, they ensure versatile use of the interior. The trayslides are U-shaped

to make the grid shelves non-tilting.

Maximum temperature stability and temperature consistency

Maximum temperature stability is ensured by the high-performance forced-air cooling system with

double ventilation, in conjunction with optimal routing of the air fl ow. The very short defrost cycles

allow the internal temperature to be kept almost constant during the defrost phase. All the laboratory

appliances are tested according to the NF X 15-140 standard with regard to maximum temperature

stability and optimal temperature consistency.

Access port for an external temperature sensor

In the laboratory appliances with Profi electronic controllers there is provision for an independent

temperature sensor allowing, for instance, a PT 100 probe or similar measuring elements to be

positioned in the interior. The diameter of the access port is 7.6 mm.

Refrigeration components

The refrigeration components are securely integrated in the ceiling area, allowing ready access. The

compressor compartment panel can be simply raised for cleaning or servicing purposes and can be fully

detached from an opening angle of 45° to provide very good access to the refrigeration components.

Electric defrosting

Hot gas defrosting

08-09_LaborSpecial_EN.indd 8

11.02.14 12:02

Castors

The LKPv and LGPv ranges come with castors as standard to allow easy and convenient cleaning

underneath and versatile use of the appliances in different rooms. The brake of the two front castors

provides a fi rm hold and ensures that the appliances cannot tip even if the grid shelves are extended.

The castors can be removed in order to transport the cabinets. Doing so the height of the cabinets will

be reduced by 13 cm.

Practical door

The doors remain open at 90° and are self-closing at less than 60° for convenient use. The integrated

lock is exceedingly sturdy and protects the stored products against unauthorised access. The door

hinges are interchangeable and the door seals are replaceable.

Infrared key with
documentation software

The alarm events and temperature

profi le data can be read out using the

infrared key available as an accessory.

Corresponding documentation software

is included enabling visualisation of the

read-out data on a PC.

Interface converter with
documentation software

A special interface converter including

documentation software is available

for central documentation of the

temperature profi le data and alarm

events of multiple appliances via the

RS 485 serial interface. Capability

is provided for interconnecting as

many as 20 laboratory appliances

and centrally documenting their

parameters.

NTC product temperature
probe

An NTC product temperature probe

for logging product temperatures

is available as a retrofi t kit for the

laboratory appliances with Profi

electronic controllers. The logged

product temperatures can either

be read by the electronic controller

or transferred to an external

documentation system via the existing

RS 485 interface.

Functional and practical accessories

U-shaped trayslides and
grid shelves

Additional U-shaped trayslides and

plastic-coated grid shelves can be

retrofi tted as required to provide extra

storage options. The robust grid

shelves have an extremely high

loading capacity of up to 60 kg.