JULABO HT30-M1 High Temperature Circulator User Manual

Page 35

Advertising
background image

HT30-M1

35

7.4. Setting the safety temperature (with shutdown function)



!

+

A L A R M !
ExtAct: xxx.xx°C

Temp/level alarm

(excess temperature protection)

• Press the key to indicate the safety temperature value on the MULTI-

DISPLAY and using a screwdriver simultaneously turn the setting
screw to the desired value
(example: 250 °C).


Setting range: 0 °C to 420 °C

in 2 °C steps



Safety device in the heating zone:
The safety device works independently from the control circlee and
supervises the temperature in the heating zone.
If the temperature of the bath liquid reaches the safety temperature, a
complete shutdown of the controlled heating device is effected.
The alarm is indicated by optical and audible signals (continuous tone).




On the MULTI-DISPLAY (LED) and DIALOG-DISPLAY (LCD) appears
the error message "E 14".







Cancel the alarm state (see page 57)

Recommendation:

• Set the safety temperature at 20 °C above the working temperature

setpoint.

• From time to time the safety device has to be checked.

With a screw driver turn back the adjustable overtemperature
protection until the shut-down point (actual temperature).

Safety device in the cooling zone:

See „Important notice“ page 25

Advertising